Tip-sample relaxation as a source of uncertainty in nanoscale scanning probe microscopy measurements

Warning

This publication doesn't include Faculty of Economics and Administration. It includes Faculty of Science. Official publication website can be found on muni.cz.
Authors

CAMPBELLOVÁ Anna KLAPETEK Petr VALTR Miroslav

Year of publication 2009
Type Article in Periodical
Magazine / Source Measurement Science and Technology
MU Faculty or unit

Faculty of Science

Citation
Field Solid matter physics and magnetism
Keywords atomic force microscopy; uncertainty; DFT; carbon nanotube
Description Nanoscale dimensional measurements are very often focused on small objects formed by only a few atomic layers in one or more dimensions. The classical convolution approach to tip-sample artefacts cannot be valid for these specimens due to the quantum-mechanical nature of small objects. As interatomic forces act on the sample and the tip of the microscope, the atoms of both relax in order to reach equilibrium positions. This leads to changes in those quantities that are finally interpreted as the atomic force microscope (AFM) tip position and influences the resultant dimensional measurements. In this paper, sources of uncertainty connected with tip-surface relaxation at the atomic level are discussed. Results of density functional theory modeling (using the tight-binding approximation software Fireball) of AFM scans on typical systems used in nanometrology, e.g., fullerenes and carbon nanotubes, on highly oriented pyrolytic graphite substrates are presented.
Related projects:

You are running an old browser version. We recommend updating your browser to its latest version.