Low-cost 3D-printed fluorescence detector for capillary electrophoresis

Investor logo
Investor logo

Warning

This publication doesn't include Faculty of Economics and Administration. It includes Faculty of Science. Official publication website can be found on muni.cz.
Authors

PŘIKRYL Jan FORET František

Year of publication 2013
Type Article in Periodical
Magazine / Source CHEMICKE LISTY
MU Faculty or unit

Faculty of Science

Citation
Field Analytic chemistry
Keywords flourescence detector; capillary electrophoresis; 3D printing
Description We have verified the possibility of Fused Deposition Modeling (type of additive manufacturing) for lab-made construction of inexpensive fluorescence detector for capillary electrophoresis using commercially available optical components. Sensitivity of fabricated and assembled detection system was tested by fluorescein solutions. Limits of detection were determined for several tested arrangements.
Related projects:

You are running an old browser version. We recommend updating your browser to its latest version.