Optimization of non-aqueous media for capillary electrophoresis of dextromethorphan and its metabolites

Investor logo

Warning

This publication doesn't include Faculty of Economics and Administration. It includes Faculty of Science. Official publication website can be found on muni.cz.
Authors

ZEISBERGEROVÁ Marta CVINGRÁFOVÁ Eliška GLATZ Zdeněk

Year of publication 2011
Type Conference abstract
MU Faculty or unit

Faculty of Science

Citation
Description Non-aqueous electrophoresis (NACE) was found to be a good alternative for the analysis of pharmaceutical drugs. In our work we have targeted the separation of dextromethorphan (DEX) and its metabolites. DEX is primarily used as an over-counter antitusive drug for temporary relief of cough. Our study has been focused on the influence of non-aqueous BGEs on the separation of DEX and its metabolites. Three organic solvents of different characteristics have been chosen for BGEs containing ammonium acetate; namely methanol, acetonitrile and 2-propanol. Separations of the object analytes with the addition of levallorphan (internal standard) were performed in a 75 um I.D. fused-silica capillary, 48.5/40 cm Ltot/Leff, respectively. The evaluation was based on the resolution of all analytes, the efficiency of levallorphan separation, electroosmotic mobility and the total analysis time.
Related projects:

You are running an old browser version. We recommend updating your browser to its latest version.